Nominal Range: 0-10 000 ppm
Maximum Overload: 20 000 ppm
Expected Operating Life: Two years in air
Output Signal: 0.003 ± 0.001 μA/ppm
Resolution: 10 ppm
Temperature Range: -20°C to +50°C
Pressure Range: Atmospheric ± 10%
Pressure Coefficient: 0.006 % signal/mBar
T90 Response Time: <110 seconds
Relative Humidity Range: 15 to 90% non-condensing
Typical Baseline Range (pure air): +25 to -150 ppm equivalent
Maximum Zero Shift (+20°C to +40°C): -150 ppm equivalent
Long Term Output Drift: <2% signal loss/month
Recommended Load Resistor: 10 Ω
Bias Voltage: Not required
Repeatability: 2% of signal
Output Linearity: Linear